Coating thickness gauge TIME®2510E with cable probe
TIME®2510E: Coating thickness gauge for Fe and NFe with permanently attached cable probe

TIME®2510E

Coating thickness measurement on ferrous and non-ferrous metals with cable probe and automatic substrate recognition

The coating thickness gauge TIME®2510E measures coatings non-destructively and works with two measuring principles: using magnetic induction (probe F) it measures non-magnetic coatings on magnetic base materials, using eddy current (probe N) it measures insulating coatings on non-magnetic metals. The base material is recognised automatically.

The probe is connected to the gauge by a permanently attached cable (no plug connector). The gauge is suitable for testing painted surfaces, automotive paint inspection, testing of coated materials and on-site quality control.

Properties of the TIME®2510E

  • two measuring principles in one gauge: magnetic induction (F) and eddy current (N)
  • cable probe permanently attached (no plug connector)
  • automatic recognition of the substrate
  • zero-point and two-point calibration to correct the system error of the probe
  • two working modes: DIRECT and BATCH, two measuring ways: CONTINUE and SINGLE
  • statistics: mean, maximum, minimum, number of readings and standard deviation
  • memory for 600 readings
  • deletion of current data, calibration data, limit data or all stored data
  • low battery indication and error alarm
  • buzzer signal during operation
  • automatic or manual shutdown

Measuring principles of the TIME®2510E

  • Magnetic induction (F): non-magnetic coating (e.g. paint, plastic, enamel, copper, zinc, aluminium, chrome) on magnetic base material (e.g. iron, steel, nickel)
  • Eddy current (N): insulating coating (e.g. enamel, rubber, paint, plastic) on non-magnetic metal (e.g. copper, aluminium, zinc, tin)
Standard delivery scope of the TIME®2510E

The TIME®2510E is supplied with:

  • Main unit TIME®2510E with permanently attached cable probe
  • Substrate (2 pcs)
  • Calibration foil
  • Instruction manual
Special accessories
  • Connecting cable
  • Dataview software
Downloads
Specifications of the TIME®2510E

Technical data of the TIME®2510E at a glance:

Parameter Probe FProbe N
Measuring principle Magnetic inductionEddy current
Probe connection Permanently attached cable probe (no plug connector)Permanently attached cable probe (no plug connector)
Measuring range 0 - 1250 µm0 - 1250 µm
0 - 40 µm (chrome plate on copper)
Display resolution 0.1 µm0.1 µm
Tolerance zero-point calibration ±(3 % H + 1) µm±(3 % H + 1.5) µm
Tolerance two-point calibration ±[(1 % to 3 %) H + 1] µm±[(1 % to 3 %) H + 1.5] µm
Min. radius of curvature (convex) 1.5 mm3 mm
Min. measuring area diameter Ø 7 mmØ 5 mm
Critical substrate thickness 0.5 mm0.3 mm
Working modes / measuring ways DIRECT, BATCH / CONTINUE, SINGLEDIRECT, BATCH / CONTINUE, SINGLE
Memory 600 readings600 readings
Power supply 2 × AAA 1.5 V batteries2 × AAA 1.5 V batteries
Working environment 0 - 40 °C, humidity 20 - 90 %, no strong magnetic field0 - 40 °C, humidity 20 - 90 %, no strong magnetic field
Dimensions 110 × 50 × 23 mm110 × 50 × 23 mm
Weight 100 g100 g

Request a quote or further information here!

Areas of application of the TIME®2510E

  • In paint shops
  • For paint and plastic coatings
  • For vehicle appraisers and vehicle insurers
  • In the automotive industry
  • In workshops
  • In metal processing
  • In the chemical industry
  • In quality control
  • In goods-in inspection
  • For on-site measurement on construction sites