Coating thickness gauge TIME®2600
TIME®2600: Coating thickness gauge for Fe and NFe with integrated printer

TIME®2600

Coating thickness measurement on ferrous and non-ferrous metals – with integrated printer

The TIME®2600 coating thickness gauge measures paints and coatings non-destructively using the magnetic induction method (non-magnetic coatings on iron and steel) and the eddy current method (non-conductive coatings on non-ferrous metals). Six different probe types are available for different applications.

The integrated printer outputs readings and statistics directly – suitable for laboratory and field use.

Features of the TIME®2600

  • two measuring principles: magnetic induction (Fe) and eddy current (NFe)
  • 6 probe types for different applications
  • two working modes: DIRECT and BATCH, two measuring ways: CONTINUE and SINGLE
  • statistics: mean, maximum, minimum, number of readings and standard deviation
  • memory for 640 readings
  • two calibration methods (zero and foil calibration)
  • integrated printer for the statistical values
  • adjustable limits with alarm
  • low battery indication and error alarm
  • backlight for the screen
  • manual or automatic shutdown
  • conforms to DIN, ISO, ASTM and BS
Standard delivery scope
  • Main unit TIME®2600
  • Probe (F1 or N1)
  • Substrate
  • Calibration foil
  • Charger
  • Printing paper
  • Instruction manual
Special accessories
  • Additional probes for special applications
Downloads
Data sheet (PDF)

pdf - 254 KB

Specifications
ParameterTIME®2600
Measuring principleMagnetic induction (F) and eddy current (N), depending on probe
Measuring rangeStandard probes F1 / N1: 0 – 1250 µm; further probes see probe table
Resolution0.1 µm (CN02: 1 µm)
Accuracy, one-point calibrationF: ±(3 % H + 1) µm; N: ±(3 % H + 1.5) µm (H = reading)
Accuracy, two-point calibrationF: ±[(1–3) % H + 1] µm; N: ±[(1–3) % H + 1.5] µm
StandardsDIN, ISO, ASTM, BS
CalibrationZero and foil calibration
Working modes / measuring waysDIRECT, BATCH / CONTINUE, SINGLE
StatisticsNumber of readings, mean, standard deviation, maximum, minimum
Memory640 readings
LimitsAdjustable, with alarm
InterfaceRS232
PrinterIntegrated printer
Power supplyNiMH rechargeable battery
Operating environment0 – 40 °C, 20 – 90 % RH, no strong magnetic field
Dimensions230 × 86 × 47 mm
ProbePrincipleRange (µm)Accuracy one-point (µm)Accuracy two-point (µm)Resolution (µm)Min. curvature radius (mm)Min. area (mm)Critical substrate thickness (mm)
F400Magnetic induction0 – 400±(3 % H + 1)±[(1–3) % H + 0.7]0.1convex 1Ø 30.2
F1Magnetic induction0 – 1250±(3 % H + 1)±[(1–3) % H + 1]0.1convex 1.5Ø 70.5
F1/90°Magnetic induction0 – 1250±(3 % H + 1)±[(1–3) % H + 1]0.1convex 1.5Ø 70.5
F10Magnetic induction0 – 10000–––10Ø 402
N1Eddy current0 – 1250
0 – 40 (chrome on copper)
±(3 % H + 1.5)±[(1–3) % H + 1.5]0.1convex 3Ø 50.3
CN02Eddy current10 – 200±(3 % H + 1)–1flatØ 7unlimited

 

Probe selection guide

Substrate / CoatingNon-magnetic coatings
(organic materials like paint, enamel, plastic)
Non-magnetic metal coatings
(chromium, zinc, copper, tin, silver, etc.)
 Coating thickness less than 100 µmCoating thickness more than 100 µmCoating thickness less than 100 µmCoating thickness more than 100 µm
Steel, iron and other magnetic metalsDiameter of testing area more than 30 mmF1: 0 – 1250 µm
F400: 0 – 400 µm
F1: 0 – 1250 µm
F10: 0 – 10 mm
F1: 0 – 1250 µm
F400: 0 – 400 µm
F1: 0 – 1250 µm
F10: 0 – 10 mm
Diameter of testing area less than 30 mmF400: 0 – 400 µmF1: 0 – 1250 µm
F400: 0 – 400 µm
F400: 0 – 400 µmF1: 0 – 1250 µm
F400: 0 – 400 µm
Copper, brass, aluminum, zinc, tin and other metalsDiameter of testing area more than 5 mmN1: 0 – 1250 µmN1: 0 – 40 µm
(for chrome plate on copper only)
Non-metallic substrateDiameter of testing area more than 7 mm––CN02: 10 – 200 µm
(mainly for testing copper foil)

Request a quote or further information here!

Areas of application of the TIME®2600

  • In paint shops
  • For powder and paint coatings
  • For automotive appraisers and automotive insurance companies
  • In the automotive industry
  • In workshops
  • In quality control
  • In bridge construction
  • In shipbuilding
  • In goods receivingIn the manufacturing process
  • In the laboratory