Universal hardenss tester HardyTest D200 with integrated impact device type D
Quick and easy
The hardness testers HardyTest D200 is especially suitable for quick and easy hardness measurements.
It is applicable with almost all hardness measurements on ferrous and non- ferrous metals. Costumers of the product include the power industry, petrochemical industry, apparatus-, automotive- and mechanical-engineering, quality control, steelworks and many more. The possibility of connecting further external impact device types makes its range of application much wider.
Features and equipment
- For fast, on site hardness testing
- Integrated impact device D
- Connection of external impact device types C, DC, D+15, DL and G also possible
- All hardness units: HL (Leeb), HB (Brinell), HRB (Rockwell B), HRC (Rockwell C), HV (Vickers), HS (Shore D)
- Large measuring range: 170-960 HLD
- Storage capacity up to 600 groups (with 1 - 32 measured values per group)
- PC data transfer via USB to DataView software or MS Hyper terminal for storing and printing
- clearly readable OLED display
- High accuracy: up to ± 6 HLD (± 0.5 % based on 800HLD)
- Automatic shutdown
- Automatic impact direction detection
- Deletion of data automatically or manually
- Limit setting (min. And max . Adjustable with arrow indicate exceeded)
- Calibration on test block in HL possible
Application
Fixtures, dies, molds, large parts, narrow spaces, axles, steering knuckles, frames, etc. Furthermore, the device is often used for material identification, and for fault checking in pressure vessels and steam turbines.
Areas of application
- Production of metal components
- Mechanical engineering
- Research and development
- Solid state physics
Standard delivery range
- HardyTest D200 with impact device D
- Test block (HLD)
- Cleaning brush
- Small supporting ring
- Battery charger
- USB cable
- Software
- Instructions Manual
- Carrying case
Optional accessory
- Support rings for convex and concave surfaces
Downloads
Specifications
Alle SaluTron® Messgeräte entsprechen nationalen (DIN) und internationalen (ISO, BS, ASTM) Normen und sie besitzen das CE-Zeichen.
Technische Änderungen vorbehalten.
HardyTest D200 | integrated impact device type D |
Hardness scales | HLD (Leeb), HB (Brinell), HRB (Rockwell B), HRC (Rockwell C), HV (Vickers), HS (Shore D) |
Measuring range | 170-960 HLD |
Accuracy | within ± 6 HLD (± 0,75% based on 800HLD) |
Application | Standard impact device for most hardness testing tasks. Connection of external impact device types C, DC, D+15, DL and G also possible. |
Material | steel/cast steel, tool steel, stainless steel, grey cast iron, nodular cast iron, cast aluminium alloys, brass, bronze, copper, forged steel |
Workpiece radius (convex/concave) | Rmin = 50 mm (with support ring Rmin = 10 mm) |
Min. Workpiece weight | 2 ~ 5kg on stable support 0.05 ~ 2kg with compact coupling |
Min. Workpiece thickness coupled | 5 mm |
Min. Surface Roughness of Workpiece | 1.6 µm (Ra) |
Min. Thickness of hardened layers | 0.8 mm |
Indentation depth | 300 HV=24 µm, 600 HV=17 µm, 800 HV=10 µm |
Impact test direction | automatic or manual |
Language | english |
Functions | automatic switch-off, battery indicator, limit value setting (with arrow display when exceeded), calibration to test block in HL |
Statistic | Setting the number of measurements to calculate the average value |
Data storage | Storage, transmission or deletion of measured values manual / automatic, 600 mean values (1-32 measured values/mean value) |
PC data processing | USB cable and software |
Standard | ASTM A956, DIN 50156 |
Test work pieces and hardness units
The ranges are stipulated by the application limits of the relevant static procedure.
Test work pieces | Hardness units | |||||
---|---|---|---|---|---|---|
HLD | HRB | HRC | HB | HV | HS | |
Steel and cast steel | 300 - 900 | 38 - 100 | 20 - 68 | 81 - 654 | 81 - 955 | 32 - 100 |
Cold work tool steel | 300 - 840 | - | 20 - 67 | - | 80 - 898 | - |
Stainless steel | 300 - 800 | 46 - 101 | - | 85 - 655 | 85 - 802 | - |
Grey cast iron | 360 - 650 | - | - | 93 - 334 | - | - |
Nodular cast iron | 400 - 660 | - | - | 131 - 387 | - | - |
Cast aluminium alloys | 170 - 570 | 23 - 84 | - | 19 - 164 | - | - |
Brass | 200 - 550 | 13 - 95 | - | 40 - 173 | - | - |
Bronze | 300 - 700 | - | - | 60 - 290 | - | - |
Copper | 200 - 690 | - | - | 45 - 315 | - | - |
Accessories for hardness tester HardyTest D200
Specifications are subject to change without prior notice.
Support rings for convex and concave surfaces | ||
---|---|---|
![]() | ||
No. | Drawing | Application |
1 | ![]() | Measurements on outer cylindrical surfaces with an radius of 10-15 mm (konvex) |
2 | Measurements on outer cylindrical surfaces with an radius of 14,5-30 mm (konvex) | |
3 | Messungen an zylindrischen Außenflächen mit einem Radius von 25-50 mm (konvex) | |
4 | ![]() | Measurements on outer cylindrical surfaces with an radius of 11-13 mm (konkav) |
5 | Measurements on outer cylindrical surfaces with an radius of 12,5-17 mm (konkav) | |
6 | Measurements on outer cylindrical surfaces with an radius of 16,5-30 mm (konkav) | |
7 | ![]() | Measurements on outer cylindrical surfaces, radius adjustable 10-∞ |
Coupling paste |
---|
Used to connect light pieces on heavier substrates (e.g. testblock) to be able to measure pieces which are lighter then required. |