TIME®2190
The TIME®2190 is a portable high-performance instrument by TIME Group Inc. (Beijing, China) for non-destructive wall thickness measurement and corrosion testing. It operates on the pulse-reflection principle and enables precise measurements on technical materials such as metals, plastics, composites, fiberglass, ceramics, and glass.
Properties
- A-scan waveform display for echo analysis
- Compatible with single and dual-element transducers
- Echo-to-echo mode for true metal thickness determination
- Thru-coat technology: simultaneous measurement of metal and coating thickness
- Automatic signal gain
- B-scan single-value display
- Fast measurement mode: up to 20 readings/second
- Data storage for 500,000 measurement values and waveforms
Applications
By using wall thickness gauges, materials that are exposed to corrosion or erosion can be monitored, ensuring the safe use of, for example, a pressure vessel.
Standard delivery scope
- Main unit
- 5 MHz dual-element transducer (DK537EE)
- Couplant
- 3× AA batteries
- User manual
- Service case
Special accessories
- Communication cable with software
- Standard test blocks
- Additional transducers for special applications
Downloads
pdf - 170 KB
Specifications
| Measuring range | 0.20 – 508.00 mm (transducer dependent) |
| Sound velocity range | 508 – 18,699 m/s |
| Display resolution | 0.001 mm / 0.01 mm / 0.1 mm selectable |
| Display | Color TFT-LCD, 320 × 240 pixels |
| Pulse generator | Adjustable square wave pulse generator |
| Pulse voltage | 60 / 110 / 150 / 200 V (selectable) |
| Gain range | 0 – 99 dB (1 dB steps) |
| Frequency range | 0.5 – 20 MHz |
| Measurement rate | Standard: 4 Hz / Fast: 20 Hz |
| Transducer profiles | 10 preset + 22 user-defined |
| Data storage | 500 files × 1,000 readings + waveforms (500,000 values total) |
| Operating temperature | 0 – 40°C |
| Power supply | 3 × AA batteries or NiMH rechargeable batteries |
| Dimensions | 187 × 87 × 43 mm |
| Weight | 360 g |
All SaluTron® measuring instruments comply with national (DIN) and international (ISO, BS, ASTM) standards and carry the CE mark.
Technical specifications subject to change without notice.
Velocities at different materials
| Material | V (in/µs) | V (m/s) |
|---|---|---|
| Aluminum; rolled | 0.2530 | 6420 |
| Aluminum; 6061-T6 | 0.2510 | 6383 |
| Beryllium | 0.5073 | 12890 |
| Brass, 70 CU, 30 Zn | 0.1850 | 4700 |
| Copper; rolled | 0.1972 | 5010 |
| Duraluminum | 0.2487 | 6320 |
| Iron | 0.2345 | 5960 |
| Lead; rolled | 0.0771 | 1960 |
| Magnesium; drawn/annealed | 0.2270 | 5770 |
| Molybdenum | 0.2470 | 6250 |
| Monel | 0.2105 | 5350 |
| Nickel | 0.2377 | 6040 |
| Steel; normal | 0.2331 | 5920 |
| Steel; lightly alloyed | 0.2259 | 5734 |
| Stainless Steel | 0.2278 | 5790 |
| Titanium | 0.2402 | 6100 |
| Tungsten, drawn | 0.2129 | 5410 |
| Uranium | 0.1330 | 3370 |
| Zinc; rolled | 0.1657 | 4210 |
| Zinc; sprayed | 0.1756 | 4460 |
| Aluminum, AL203 99.5% | 0.4013 | 10190 |
| Quartz Glass | 0.2349 | 5968 |
| Melted | 0.2335 | 5932 |
| Pyrex Glass | 0.2220 | 5640 |
| Lucite | 0.1055 | 2680 |
| Plexiglass | 0.1077 | 2735 |
| Nylon | 0.1031 | 2620 |
| Polyethylene | 0.0705 | 1950 |
| Polystyrene | 0.0925 | 2350 |
| Silicone Rubber | 0.0373 | 948 |
| Water | 0.0590 | 1490 |
Overview of compatible transducers for the ultrasonic wall thickness gauge TIME®2190
| Transducer Type | Measuring Range Steel (mm) | Frequency | Drawing | Note |
|---|---|---|---|---|
| DK537EE | 1 – 508 | 5.0 MHz Dual-Element | ![]() | Standard delivery |
| DC250 | 2.5 – 508 | 2 MHz Dual-Element | ![]() | Application: Measuring 225 mm with coating |
| DLK1025 | 0.5 – 12.7 | 10 MHz Single-Element with delay line | ![]() | Application: Measuring 0.8 mm with coating |
| DCC110 | 7.62 – 508 | 1 MHz Single-Element | ![]() | Measuring FRP materials |
| DLK1225 | 0.5 – 5.08 | 12 MHz Single-Element | ![]() | Measuring very small workpieces |
| DHT-537 | 1 – 508 | 5 MHz Dual-Element | ![]() | High temperature up to 500°C |
| DEFM2-SE | 0.15 – 20 | 15 MHz Single-Element | ![]() | Measuring thin workpieces |
| SZ2.5P | 3 – 300 | 2.5 MHz Dual-Element | ![]() | |
| TSTU32 | 5 – 40 (gray iron) | 2 MHz Single-Element | ![]() | Gray cast iron |
| ZW5P | 4 – 80 | 5 MHz Dual-Element | ![]() | High temperature up to 300°C |
Areas of application of wall thickness gauges
- Steel plates
- Sheet metal
- Finishing parts
- Pipes
- Pressure vessels
- For example in the petroleum, chemical, metal, ship, space, and aerospace industries











![[Translate to English:] [Translate to English:] Shore Durometer](/fileadmin/_processed_/5/7/csm_haertepruefgeraet-shore-durometer_cf6f60ff1c.webp)




![Wall Thickness Gauge TIME®2190 [Translate to English:]](/fileadmin/_processed_/5/d/csm_time-2190_3d70bc3a68.webp)
![[Translate to English:] [Translate to English:]](/fileadmin/_processed_/5/1/csm_time3100_4982bd6c41.webp)

![[Translate to English:] [Translate to English:]](/fileadmin/_processed_/9/1/csm_rauheitsmessgeraet-tr-110-time-3110_f4e53f7305.webp)
![[Translate to English:] [Translate to English:]](/fileadmin/_processed_/2/5/csm_rauheitsmessgeraet-tr200-time3200-tr220-time3202_7e08574567.webp)
![[Translate to English:] [Translate to English:]](/fileadmin/_processed_/2/1/csm_glanzmessgeraet-glosstector_30b19a3b3a.webp)